Octiv Suite Specifications

Measuring Parameters (Range)

Voltage Range 20 – 3000 Vrms
Current Range 0.1 – 20 Arms
Phase Range ± 90º
Harmonic (Voltage, Current and Phase) Up to 15 Harmonics
Frequency Range 350 kHz - 300 MHz
Fundamental Frequencies 5 Simultaneous
Impedance 1 to 500Ω
Power Real, Forward and Reflected (Watt) 200mW to 12KW
Power Real, Forward and Reflected (dBm) 25dBm to 70dBm

Measuring Plasma Parameters

Ion Flux (based on 300mm electrode) 1 A/m² to 100 A/m²
Plasma Resistance 1 to 500Ω
Non Linear Sheath Impedance .1 to 500Ω

Pulsed Parameters (Time)

Voltage Time 1μs
Current Time 1μs
Phase Time 1μs
Harmonic (Voltage, Current and Phase) Time 1μs
Frequency Time 1μs
Impedance Time 1μs
Power Real, Forward and Reflected (Watt) Time 1μs
Power Real, Forward and Reflected (dBm) Time 1μs

Measuring Parameters (Accuracy)

Voltage Accuracy ± 1%
Current Accuracy ± 1%
Phase Accuracy ± 1º
Harmonic (Voltage, Current and Phase) Accuracy ± 5%
Frequency Accuracy ± 10kHz
Impedance ± 1%
Power Real, Forward and Reflected (Watt) ± 1%
Power Real, Forward and Reflected (dBm) ± 1%

Measuring Parameters (Resolution)

Voltage Resolution 0.25V
Current Resolution 10mA
Phase Resolution 0.01°
Harmonic Voltage Resolution 0.25V
Harmonic Current Resolution 10mA
Harmonic Phase Resolution 0.01°
Frequency Resolution 1kHz
Impedance Resolution ± 1%
Power Real, Forward and Reflected (Watt) Resolution ± 1%
Power Real, Forward and Reflected (dBm) Resolution ± 1%

Sensor Specifications

Connectors BNC-Female, BNC-Male, HN-Female, HN-Male, LC-Female, LC-Male, N-Female, N-Male, SMA-Female, SMA-Male, 7/16 Jack, IEC Type 169-4, 7/16 Plug, IEC Type 169-4, Mini UHF-Female, UHF-Female, UHF-Male, 1-5/8" EIA Fixed, 7/8" EIA, TNC-Female, TNC-Male, and Open Term. #10-32 Nut
Dimensions Standard unit: 70mm x 70mm x 55mm | Industrial unit: 70mm x 107mm x 55 | Custom designs upon request
Number of fundamentals (F0) Maximum of 5 simultaneously
RF Power Max 12.5 kW (limited by connector)
Operating Temperature 0 to +40°C (32 to 104°F)
Storage Temperature -20 to +80°C (-4 to +176°F)
Uniformity 2% Maximum
Harmonic Content Measured (No Limit within Range)
Connectors All Standard Connectors Available
Sensor Impedance 50Ω
Certification CE mark
Calibration Cycle 12 Months

Application Software

Operating System Windows 2000 / XP / Vista / Windows 7 / Windows 8 / Windows 10

Operating Parameters

Impedance 0Ω to 5,000Ω
Pulsed Repetition Frequency 10Hz to 100KHz
Voltage 20V to 3,000V
Current 0.1A to 100A
Phase ±90º, ±180º
Power Frequency MF (350kHz to 1MHz) • RF (1MHz to 100MHz)